Description
Technical Specifications:
-
Freezing chamber temperature range: between -10 C to -30 C.
-
Cooling down time to -30 C: 60 min.
-
Freeze shelf minimum temperature: -45 C.
-
Lowest temperature of peltier unit on freeze shelf: -55 C.
-
Peltier working time: 15 minutes.
-
Maximum specimen size: 35mm x35mm.
-
Vertical stroke: 60mm, Object feed: 20mm.
-
Coarse feeding speed: 0.7mm/s; 0.35mm/s.
-
Section thickness range: 1μm-80μm adjustable.
-
1μm-20μm, in 1μm increment
-
20μm-40μm, in 2μm increments
-
40μm-80μm, in 5μm increments
-
-
Trimming thickness range: 10μm-400μm, adjustable.
-
10μm-50μm, in 5μm increments
-
50μm-100μm, in 10μm increments
-
100μm-400μm, in 50μm increments
-
-
Retraction range: 0-80μm adjustable; in 5μm increment.
-
Voltage and frequency: AC220V~10% 50HZ(Standard) Special requirement (AC110V) may customize by order.
-
Power : 850VA.
-
Dimension(WX DX H): 660x640x1130mm.
-
Net weight: 130kg.
Features:
-
It is designed under overall engineering and ergonomic concept. All the components are processed with computer numerical controlled machine.
-
Specimen retraction function, protects specimen from blade injury.
-
Trimming/Cutting button makes it easy to switch from trim and section mode.
-
Equip with counter to calculate total pieces and thickness.
-
Dual compressors, which are made in Germany, freeze the Cryostat chamber, freezing shelf, knife holder and specimen Clamp separate with pro-environment refrigerant R404.
-
Temperature can drop to working condition in 60 minutes after switching on.
-
Adopt to UVC sterilize for 35 minuts every time.
-
Peltier may be on or off manually.
-
Heated, removable sliding window.
-
Two defrost methods: definite time and manual.
-
The high-precision microtome is enclosed outside the cryostat chamber to protect it from thermal expansion and contraction and keep minimum service and maintenance.
-
Self-explanatory symbols for all functions and displays, easy to learn and operate.
-
LCD screen can display date such as sample travel, thickness, function controlling ,day, time, temperature,
timed on/off and so on. -
Stand by mode, after initializing, the temperature in cryostat chamber can be controlled between -4 C and -9 C by computer automatically; when stop initializing, the temperature in freeze chamber will arrive to working temperature within 10 minutes.
-
Control panel with locking function to avoid wrong operation.
-
Handwheel locking.
-
Sensor self-diagnosis function.
-
Wide freezing shelf may put 10 specimen cassettes.
-